El HC2S3 es un sensor preciso y robusto para medir la temperatura y humedad relativa del aire, ideal para aplicaciones desatendidas de larga duración. Para medir la humedad relativa, incluye un sensor capacitivo avanzado. Para garantizar fiabilidad y funcionamiento, incluye filtro para protección contra el polvo y partículas.
Leer másEl HC2S3 incluye un filtro de polietileno que protege el sensor del polvo fino y partículas, minimizando la absorción y retención de agua. Opcionalmente está disponible un filtro de teflon adecuado para ambientes marinos. El tiempo de respuesta es más lento si se utiliza el filtro de teflon.
Número de FAQs relacionadas con HC2S3: 9
Expandir todoDesplegar todo
It is possible that an older version of Short Cut is being used. Download the latest version of Short Cut.
Campbell Scientific offers two filter tips:
The sensor should not be cleaned with water. If the sensor is dirty, the filter should be changed or cleaned more frequently. A dirty filter can be cleaned with distilled water.
To measure ambient temperature and relative humidity (RH), mount the HC2S3-L probe inside a 41003-5 or 41003-5A 10-Plate Radiation Shield. The 27731 hex plug is shipped with the HC2S3-L so that it can be used with older 10-plate radiation shield models. The 6-plate radiation shield (pn 41303-5A) cannot be used with the HC2S3-L.
Most Campbell Scientific sensors are available as an –L, which indicates a user-specified cable length. If a sensor is listed as an –LX model (where “X” is some other character), that sensor’s cable has a user-specified length, but it terminates with a specific connector for a unique system:
If a sensor does not have an –L or other –LX designation after the main model number, the sensor has a set cable length. The cable length is listed at the end of the Description field in the product’s Ordering information. For example, the 034B-ET model has a description of “Met One Wind Set for ET Station, 67 inch Cable.” Products with a set cable length terminate, as a default, with pigtails.
If a cable terminates with a special connector for a unique system, the end of the model number designates which system. For example, the 034B-ET model designates the sensor as a 034B for an ET107 system.
Not every sensor has different cable termination options. The options available for a particular sensor can be checked by looking in two places in the Ordering information area of the sensor product page:
If a sensor is offered in an –ET, –ETM, –LC, –LQ, or –QD version, that option’s availability is reflected in the sensor model number. For example, the 034B is offered as the 034B-ET, 034B-ETM, 034B-LC, 034B-LQ, and 034B-QD.
All of the other cable termination options, if available, are listed on the Ordering information area of the sensor product page under “Cable Termination Options.” For example, the 034B-L Wind Set is offered with the –CWS, –PT, and –PW options, as shown in the Ordering information area of the 034B-L product page.
Note: As newer products are added to our inventory, typically, we will list multiple cable termination options under a single sensor model rather than creating multiple model numbers. For example, the HC2S3-L has a –C cable termination option for connecting it to a CS110 instead of offering an HC2S3-LC model.
Note the difference between calibration and a field check. Calibration cannot be done in the field, as it requires an experienced technician and specialized equipment.
Field checks of measurements can be done to determine if the data make sense with the real-world conditions. Follow these steps to field check a sensor:
Many Campbell Scientific sensors are available with different cable termination options. These options include the following:
Note: The availability of cable termination options varies by sensor. For example, sensors may have none, two, or several options to choose from. If a desired option is not listed for a specific sensor, contact Campbell Scientific for assistance.
To incorporate a sensor that is compatible with wireless sensor interfaces into a wireless network, a CWS900-series wireless sensor interface is needed, as well as an A205 CWS-to-PC interface to configure it.
Electronics Operating Limits | -40° to +100°C |
Rango temperatura almacenamiento | -50° to +100°C |
Descripción filtro | Polyethylene (standard) or Teflon (optional, ordered separately) |
Consumo |
|
Voltaje alimentación | 5 to 24 Vdc |
Tiempo caldeo |
1.5 s (typical) The startup time is a Rotronics specification. For programming, Campbell Scientific recommends a 2 s delay at 60°C, a 3 s delay at 0°C, and a 4 s delay at -40°C. |
Maximum Startup Current | < 50 mA (for 2 μs) |
Analog Outputs |
|
Diameter | 15 mm (0.6 in.) |
Longitud |
|
Peso | 10 g (0.35 oz) |
Temperatura del aire |
|
Sensor | PT100 RTD, IEC 751 1/3 Class B |
Rango de medida |
-40° to +60°C (default) Air temperature measurement ranges from -50° to +100°C are possible; contact Campbell Scientific for more information. |
Rango señal de salida | 0 to 1 V |
Precisión | ±0.1°C with standard configuration settings (at 23°C) |
Estabilidad a largo plazo | < 0.1°C/year |
Sensor Time Constant - Standard PE Filter | ≤ 22 s (63% step change [1 m/s air flow at sensor]) |
Sensor Time Constant - Optional Teflon Filter | ≤ 30 s (Typical 4 s, 63% of a step change [1 m/s air flow at sensor]) |
Temperature Accuracy | Refer to graph in probe manual. |
Humedad relativa |
|
Sensor | ROTRONIC® Hygromer IN-1 |
Rango de medida | 0 to 100% RH (non-condensing) |
Rango señal de salida | 0 to 1 Vdc |
Estabilidad a largo plazo | < 1% RH per year |
Precisión | ±0.8% RH with standard configuration settings (at 23°C) |
Sensor Time Constant - Standard PE Filter | ≤ 22 s (63% of a 35 to 80% RH step change [1 m/s air flow at sensor]) |
Sensor Time Constant - Optional Teflon Filter | ≤ 30 s (Typical 10 s, 63% of a 35 to 80% RH step change [1 m/s air flow at sensor]) |
RH Accuracy over Temperature | Refer to graph in probe manual. |
Nota: lo siguiente muestra información de compatibilidad notable. No es una lista de todos los productos compatibles.
Producto | Compatible | Nota |
---|---|---|
CR1000 (retired) | ||
CR1000X | ||
CR200X (retired) | ||
CR216X (retired) | ||
CR3000 (retired) | ||
CR5000 (retired) | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000X (retired) |
Producto | Compatible | Nota |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR10X (retired) | ||
CR23X (retired) | ||
CR500 (retired) | ||
CR510 (retired) | ||
CR9000 (retired) |
El HC2S3 debe montarse dentro de los protectores de radiación con ventilación natural, que incluyen los elementos para fijación a mástil o poste vertical.