Disponibilidad limitada ›
Este producto sufre disponibilidad limitada. En caso de que quedara descatalogado, podrían seguir estando disponibles accesorios, recambios, u otros servicios.
CR9058E 60 V Isolated Analog Input Module with CR9058EC

Resumen

The CR9058E is a ±60-volt input module for the CR9000X and CR9000 measurement and control systems. The CR9058E has 10 differential inputs. Each channel has its own isolation ground, which can be used to connect a shielded cable. The CR9058E has an on-board resistance temperature detector (RTD) to provide a reference temperature measurement for thermocouple measurements. It ships with one CR9058EC Easy Connector Module.

Leer más

Ventajas y características

  • CR9058EC disconnects sensors easily so CR9000X can be moved from location to location
  • Can be added to or removed from CR9000X to support changing measurement needs

Imágenes

Descripción detallada

The CR9058E is an I/O module for the CR9000X-series and CR9000-series dataloggers. It contains 10 isolated differential channels and an internal RTD. The differential channels are used to measure thermocouples or other devices that have a low level voltage output. Each channel has its own isolated ground for connecting a shielded cable, and its own 24-bit A/D converter that supplies input isolation for up to ±60 Vdc. An internal digital signal processor filters noise. The module's RTD typically provides the reference temperature for thermocouple measurements.

The CR9058E includes one CR9058EC easy connector module. Easy connector modules allow sensor wiring to remain connected while the input module’s measurement electronics and the rest of the datalogger system are used elsewhere. Additional CR9055ECs are purchased as a separate line item.


Especificaciones

-NOTA- Additional specifications are listed in the CR9000X(C) Specifications Sheet.
Analog Channels 10 isolated differential
Analog Voltage Range ±60000 mV
Temperatura funcionamiento -40° to +70°C
Dimensiones
  • 24.1 x 2.5 x 11.5 cm (9.5 x 1.0 x 4.5 in.) without EC
  • 24.1 x 2.5 x 16.3 cm (9.5 x 1.0 x 6.4 in.) with EC
Peso
  • 540 g (19.1 oz) for card and EC
  • 357 g (12.6 oz) for card only
  • 183 g (6.5 oz) for EC only

Gain Error

-40° to +50°C ±0.02% of reading
-40° to +70°C ±0.07% of reading

Offset Error

-40° to +50°C ±0.01% of FSR
-40° to +70°C ±0.01% of FSR

Measurement Resolution

±2 Vdc Range ±10 μV
20 Vdc Range ±100 μV
±60 Vdc Range ±300 μV

Typical Current Drain

Operating 360 mA
Reposo 5 mA

Compatibilidad

Nota: lo siguiente muestra información de compatibilidad notable. No es una lista de todos los productos compatibles.

Dataloggers

Producto Compatible Nota
CR1000 (retired)
CR200X (retired)
CR216X (retired)
CR3000 (retired)
CR5000 (retired)
CR800 (retired)
CR850 (retired)
CR9000X (retired)

Miscelaneo

Producto Compatible Nota
21X (retired)
CR10 (retired)
CR10X (retired)
CR23X (retired)
CR500 (retired)
CR510 (retired)
CR9000 (retired)

Figura en